ProRes scientific understands the importance of your Glove box, so we as a team having experience over a decade have succeeded in providing service for any existing Glove box without any exception.

  • Inspect Gloves/Glove ports/Front Panel Windows
  • Feedthroughs and KF Fittings
  • Working Gas Supply Connections
  • Vacuum System Connections
  • Replacement of HEPA Filters
  • Replace Small Antechamber O-Rings
  • Replace Large Antechamber O-Rings
  • Vacuum Pump Oil replacement
  • Replacement of Copper Catalyst and Molecular Sieves for Purifier Column
  • Replacement of Activated Carbon for Organic Vapor Absorber
  • Replacement of Activated Carbon/Acidic Vapor Cartridges where fitted.
  • Cleaning of Moisture Probes

Glovebox Maintenance

  • Periodic maintenance of your glove box system is essential to achieve the atmosphere quality that is required for your air-free process or application. Manufacturers state that inert gas atmospheres of <1ppm Oxygen and Water as Moisture are achievable, but this depends on the glove box system working to its full potential.
  • Regular maintenance of important glove box components including Vacuum System, Purification Column media, Solvent Absorption media and Seals/O-rings are essential to keeping your inert atmosphere within specification, and allowing the continuing purification of your inert gas.
  • Regular calibration/cleaning of Oxygen and Moisture sensors are also crucial in providing the user with accurate qualitative readings, giving confidence that your glove box atmosphere is satisfactory for you to perform your process within safety and quality tolerances